B-1(+) shimming with SAR reduction in high-field MRI

B1+ shimming with SAR reduction in highfield
  • Son, H. W.; 
  • Cho, Y. K.; 
  • Gopinath, A.; 
  • Vaughan, J. T.; 
  • Lee, C. H.; 
  • ... Yoo, H.
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

In high-field magnetic resonance imaging systems, parallel excitation with a multi-channel transmitter and receiver has been widely investigated to mitigate inhomogeneity. Multi-channel RF coils employed in the parallel imaging may alleviate the non-uniformity in the field and each element can be independently controlled by adjusting the phase and amplitude of the excitation. In this paper, convex optimization for each coil elements to homogenize and to reduce E fields is proposed. Improved homogeneity and reduced specific absorption rate over the whole field of view is achieved.

제목
B-1(+) shimming with SAR reduction in high-field MRI
제목 (타언어)
B1+ shimming with SAR reduction in highfield
저자
Son, H. W.; Cho, Y. K.; Gopinath, A.; Vaughan, J. T.; Lee, C. H.; Yoo, H.
DOI
10.1080/09205071.2013.817958
발행일
2013-10
유형
정기학술지(Article(Perspective Article포함))
저널명
Journal of Electromagnetic Waves and Applications
권
27
호
12
페이지
1521 ~ 1524