Comparison of titanium hydride (TiH2) and paraffin as neutron moderator material in a prompt gamma scanning system

Citations

WEB OF SCIENCE

19
Citations

SCOPUS

20

초록

If the proton therapy modality allowing excellent dose localization to the tumor volume and negligible exit dose in the patient is to be taken advantage of, a proper dosimetry system is required. We have recently designed a prompt-gamma scanning system to measure the proton range in situ by using a Monte Carlo technique employing MCNPX, FLUKA, and Sabrina (TM). The gamma scanning system was designed to measure only the right-angled prompt gammas passing through a narrow collimation hole in order to correlate the position with the dose distribution. The collimation part of the scanning system, which has been constructed to measure the gammas at a 70-MeV proton energy, is made of a set of paraffin, berated carbon, and lead layers to shield the high-energy neutrons and secondary photons. However, in this study, we tried a titanium-hydride (TiH2) layer instead of the paraffin to reduce the background radiation. Our result shows that titanium hydride gives a better result when the distance between the phantom and the gamma scanner is short.

키워드

proton therapyprompt gammasMonte CarloneutronsdosimetryFACILITY
제목
Comparison of titanium hydride (TiH2) and paraffin as neutron moderator material in a prompt gamma scanning system
저자
Seo, Kyu SeokKim, Chan HyeongKim, Jong Won
발행일
2006-04
유형
Article; Proceedings Paper
저널명
Journal of the Korean Physical Society
48
4
페이지
855 ~ 858