자동혈압계 성능평가를 위한 인체혈압 시뮬레이터 개발

A Simulator for the Validation of Non-invasive Blood Pressure(NIBP) Monitoring Devices
  • 도일
  • 임현균
  • 안봉영
  • 지영준
  • 이종실
  • ... 오재훈

초록

Blood pressure is one of the important vital signs for monitoring the medical condition of a patient. Automated NIBP(non-invasive blood pressure) monitoring devices calculate systolic and diastolic blood pressures from the oscillation in cuff pressure caused by a pulsation of an artery. To validate the NIBP devices, we developed a simulator to supply the oscillometric waveforms obtained from human subjects. The simulator provided pressure pulses to device-under-test and device readings were compared to the auscultatory references. Fully automated simulation system including OCR(optical character recognition) were developed and used for NIBP monitoring devices. The validation results using the simulator agreed well with previous clinical validation. More validation studies using the standardized oscillometric waveforms would be required for the replacement of clinical trials to validate a new automated NIBP monitoring device.

키워드

Non-invasive blood pressure(NIBP)SimulatorOscillometric waveformClinical validationPressure Pulse
제목
자동혈압계 성능평가를 위한 인체혈압 시뮬레이터 개발
제목 (타언어)
A Simulator for the Validation of Non-invasive Blood Pressure(NIBP) Monitoring Devices
저자
도일임현균안봉영지영준이종실오재훈
DOI
10.9718/JBER.2017.38.3.111
발행일
2017-06
저널명
의공학회지
38
3
페이지
111 ~ 115