Comparison of the Bone Harvesting Capacity of an Intraoral Bone Harvesting Device and Three Different Implant Drills

  • Lim, Hyun-Chang
  • Ha, Kyung-In
  • Hong, Ji-Youn
  • Han, Ji-Young
  • Shin, Seung-Il
  • 외 3명
Citations

WEB OF SCIENCE

7
Citations

SCOPUS

7

초록

The aim of the present study was to compare bone-collecting capacity of bone harvesting device and minimally irrigated low-speed drilling using three implant systems. One bone harvesting device and three commercially available drill systems were compared using the osteotomies on bovine rib bones. The amount of the collected bone particle and particle size ([removed]1000 μm: large) were measured. Total wet (1.535±0.232 mL) and dry volume (1.147±0.425 mL) of the bone particles from bone harvesting device were significantly greater than three drill systems (wet volume: 1.225±0.187-1.27±0.29 mL and dry volume: 0.688±0.163-0.74±0.311 mL) (P[removed]1000 μm) were harvested significantly greater by bone harvesting device than minimally irrigated low-speed drilling. The composition of particle size in all harvesting methods was similar to each other.

키워드

MAXILLARY SINUS AUGMENTATIONPARTICLE-SIZEAUTOGENOUS BONEGRAFTOSTEOTOMYVOLUME
제목
Comparison of the Bone Harvesting Capacity of an Intraoral Bone Harvesting Device and Three Different Implant Drills
저자
Lim, Hyun-ChangHa, Kyung-InHong, Ji-YounHan, Ji-YoungShin, Seung-IlShin, Seung-YunHerr, YeekChung, Jong-Hyuk
DOI
10.1155/2017/7819080
발행일
2017-00
유형
Article
저널명
BioMed Research International
2017
페이지
1 ~ 6

파일 다운로드