Matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay was able to detect carbapenemase producers, including SIM-1 or OXA-51, within 4 hours using 20 mu L of 0.5 g/L ertapenem solution as a substrate. This assay is more rapid and accurate than the modified Hodge test and 3-dimensional extract bioassay. Hence, it can be used an alternative test to identify carbapenemase-mediated carbapenem resistance in Gram-negative bacteria. (C) 2013 Elsevier Inc. All rights reserved.

키워드

Acinetobacter spp; Carbapenemase; MALDI-TOF MS; OXA; SIM-1; MODIFIED HODGE TEST; RAPID DETECTION; BAUMANNII; RESISTANCE; GENE
제목
Comparison of matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay with conventional methods for detection of IMP-6, VIM-2, NDM-1, SIM-1, KPC-1, OXA-23, and OXA-51 carbapenemase-producing Acinetobacter spp., Pseudomonas aeruginosa, and Klebsiella pneumoniae
저자
Lee, Wonmok; Chung, Hae-Sun; Lee, Yang soon; Yong, Dongeun; Jeong, Seok Hoon; Lee, Kyungwon; Chong, Yunsop
DOI
10.1016/j.diagmicrobio.2013.07.005
발행일
2013-00
유형
정기학술지(Article(Perspective Article포함))
저널명
Diagnostic Microbiology and Infectious Disease
권
77
호
3
페이지
227 ~ 230