32.1 A Behind-The-Ear Patch-Type Mental Healthcare Integrated Interface with 275-Fold Input Impedance Boosting and Adaptive Multimodal Compensation Capabilities

  • Kim, Hyunjoong
  • Kim, Myeongwoo
  • Lee, Kwangmuk
  • Cho, Sanghyeon
  • Park, Chan Sam
  • ... Jang, Dong Pyo
  • 외 3명
Citations

SCOPUS

27

초록

Conventional brain activity monitoring devices including head bands and scalp caps are still not appropriate for daily-life monitoring applications because of their wearing discomfort and electrode-related artifacts. A proposed behind-the-ear (BTE) device, where motion artifacts are significantly reduced, would be a promising candidate for continuous mental healthcare platforms. This proposed BTE-based measurement allows multimodal bio-signals such as electroencephalogram (EEG), electrocardiogram (ECG), photoplethysmogram (PPG), and galvanic skin response (GSR), whereas most conventional brain devices only monitor EEG. However, these bio-signals become much weaker at the BTE location, and their signal quality is more important, being seriously degraded by artifact-related corruptions during daily activities. For this purpose, an offset-compensated auxiliary path (OCAP) and a dual-resolution external positive feedback loop (DR-EPFL) are proposed to boost the input impedance and to improve its AC characteristics further. For small-featured device sizes, the number of electrodes is reduced by utilizing a proposed GSR-embedded ECG readout structure where electrodes are shared and monitored together. For system-level feasibility, a direct-conversion PPG readout with a proposed parasitic capacitance compensation, a transcutaneous vagus-nerve stimulator (texttVNS) with an electrode-monitored adaptive charge pump, and a high-resolution bio-impedance channel are integrated together. A BTE-based mental healthcare patch-type device prototype is also manufactured and its multimodal functionality is experimentally verified.

키워드

BrainCapacitanceElectric impedanceElectric impedance measurementElectrocardiogramsElectroencephalographyElectrophysiologyFeedbackHealth careActivity monitoringBiosignalsBrain activityCompensation capabilityGalvanic skin responseInput impedanceIntegrated interfaceMulti-modalPatch typePhotoplethysmogramElectrodes
제목
32.1 A Behind-The-Ear Patch-Type Mental Healthcare Integrated Interface with 275-Fold Input Impedance Boosting and Adaptive Multimodal Compensation Capabilities
저자
Kim, HyunjoongKim, MyeongwooLee, KwangmukCho, SanghyeonPark, Chan SamSong, SolwoongKeum, Dae SikJang, Dong PyoKim, Jae Joon
DOI
10.1109/ISSCC42615.2023.10067723
발행일
2023-02
유형
Conference paper
저널명
Digest of Technical Papers - IEEE International Solid-State Circuits Conference
2023-February
페이지
478 ~ 480