Uncertainty evaluation of photoluminescence quantum yield measurement in an integrating hemisphere-based instrument

  • Yoo, Jaekak; 
  • Gene, Jinhwa; 
  • Kim, Jaeseok; 
  • Park, Seung-Nam; 
  • Lim, Seong Chu; 
  • ... Jeong, Mun Seok; 
  • 외 2명
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

We present an integrating hemisphere-based (i.e., a variant of integrating spheres) implementation of the indirect illumination method for absolute photoluminescence quantum yield measurements, which is a recommended method in the international standard IEC 62607-3-1:2014. We rigorously formulated a mathematical model and a measurement procedure for the absolute photoluminescence quantum yield measurement in the integrating hemisphere-based system. The measurement system was calibrated using an Hg-Ar discharge lamp and spectral irradiance standard lamps for wavelength and relative spectral radiant flux scales, respectively. Furthermore, we identified and evaluated uncertainty components involved in the photoluminescence quantum yield (PLQY) measurement. To validate our measurement system, we applied it to the two de facto standard dyes: quinine bisulfate (QBS) and fluorescein (FLS). Consequently, their PLQY values were determined to be 0.563±0.024 (=2) and 0.876±0.032 (=2) for, respectively, QBS and FLS, which are consistent with previous reports.

키워드

Mercury compounds; Photoluminescence; Uncertainty analysis; Quantum yield; Illumination method; Indirect illumination; Integrating spheres; International standards; Measurement procedures; Measurement system; Photoluminescence quantum yields; Spectral irradiance; Uncertainty evaluation; Yield measurement
제목
Uncertainty evaluation of photoluminescence quantum yield measurement in an integrating hemisphere-based instrument
저자
Yoo, Jaekak; Gene, Jinhwa; Kim, Jaeseok; Park, Seung-Nam; Lim, Seong Chu; Jeong, Mun Seok; Lee, Tae Geol; Park, Seongchong
DOI
10.1364/AO.490222
발행일
2023-06
유형
Article
저널명
Applied Optics
권
62
호
18
페이지
4805 ~ 4812