Multi-slit prompt-gamma camera for locating of distal dose falloff in proton therapy

Citations

WEB OF SCIENCE

21
Citations

SCOPUS

25

초록

In this research, a multi-slit prompt-gamma camera was developed to locate the distal dose falloff of the proton beam spots in spot scanning proton therapy. To see the performance of the developed camera, therapeutic proton beams were delivered to a solid plate phantom and then the prompt gammas from the phantom were measured using the camera. Our results show that the camera locates the 90% distal dose falloff (= d90%), within about 2-3 mm of error for the spots which are composed 3.8 x 10(8) protons or more. The measured location of d90% is not very sensitive to the irradiation depth of the proton beam (i.e., the depth of proton beam from the phantom surface toward which the camera is located). Considering the number of protons per spot for the most distal spots in typical treatment cases (i.e., 2 Gy dose divided in 2 fields), the camera can locate d90% only for a fraction of the spots depending on the treatment cases. However, the information of those spots is still valuable in that, in the multi-slit prompt-gamma camera, the distal dose falloff of the spots is located solely based on prompt gamma measurement, i.e., not referring to Monte Carlo simulation.

키워드

Proton therapyPrompt gammaBeam range verificationDistal dose falloffBEAM RANGE VERIFICATIONCOMPTON CAMERARADIATION HARDNESSPENCIL BEAMSSLIT CAMERATARGETSSYSTEM
제목
Multi-slit prompt-gamma camera for locating of distal dose falloff in proton therapy
저자
Park, Jong HoonKim, Sung HunKu, YoungmoKim, Chan HyeongLee, Han RimJeong, Jong HwiLee, Se ByeongShin, Dong Ho
DOI
10.1016/j.net.2019.03.008
발행일
2019-08
유형
Article
저널명
Nuclear Engineering and Technology
51
5
페이지
1406 ~ 1416

파일 다운로드