DEVELOPMENT AND EVALUATION OF A PHANTOM FOR MULTI-PURPOSE DOSIMETRY IN INTENSITY-MODULATED RADIATION THERAPY

Citations

WEB OF SCIENCE

6
Citations

SCOPUS

6

초록

A LEGO-type multi-purpose dosimetry phantom was developed for intensity-modulated radiation therapy (IMRT), which requires various types of challenging dosimetry. Polystyrene, polyethylene, polytetrafluoroethylene (PTFE), and polyurethane foam (PU-F) were selected to represent muscle, fat, bone, and lung tissue, respectively, after considering the relevant mass densities, elemental compositions, effective atomic numbers, and photon interaction coefficients. The phantom, which is composed of numerous small pieces that are similar to LEGO blocks, provides dose and dose distribution measurements in homogeneous and heterogeneous media. The phantom includes dosimeter holders for several types of dosimeters that are frequently used in IMRT dosimetry. An ion chamber and a diode detector were used to test dosimetry in heterogeneous media under radiation fields of various sizes. The data that were measured using these dosimeters were in disagreement when the field sizes were smaller than 1.5 x 1.5 cm(2) for polystyrene and PTFE, or smaller than 3 x 3 cm(2) for an air cavity. The discrepancy was as large as 41% for the air cavity when the field size was 0.7 x 0.7 cm(2), highlighting one of the challenges of IMRT small field dosimetry. The LEGO-type phantom is also very useful for two-dimensional dosimetry analysis, which elucidates the electronic dis-equilibrium phenomena on or near the heterogeneity boundaries.

키워드

IMRTPhantomDosimetryHeterogeneityIMRTCHAMBERFIELDS
제목
DEVELOPMENT AND EVALUATION OF A PHANTOM FOR MULTI-PURPOSE DOSIMETRY IN INTENSITY-MODULATED RADIATION THERAPY
저자
Jeong, Hae SunHan, YoungyihKum, OyeonKim, Chan HyeongPark, Joo Hwan
DOI
10.5516/NET.2011.43.4.399
발행일
2011-08
유형
Article
저널명
Nuclear Engineering and Technology
43
4
페이지
399 ~ 404

파일 다운로드