상세 보기
CAD 모델 기반의 4D CT 데이터 제작 의용공학 융합 프로그램 개발
- 서정민;
- 한민철;
- 이현수;
- 이세형;
- 김찬형
초록
In the present study, we developed the 4D CT data generation program from CAD-based models. To evaluate the developed program, a CAD-based respiratory motion phantom was designed using CAD software, and converted into 4D CT dataset, which include 10 phases of 3D CTs. The generated 4D CT dataset was evaluated its effectiveness and accuracy through the implementation in radiation therapy planning system (RTPS). Consequently, the results show that the generated 4D CT dataset can be successfully implemented in RTPS, and targets in all phases of 4D CT dataset were moved well according to the user parameters (10 mm) with its stationarily volume (8.8 cc). The developed program, unlike real 4D CT scanner, due to the its ability to make a gold-standard dataset without any artifacts constructed by modality’s movements, we believe that this program will be used when the motion effect is important, such as 4D radiation treatment planning and 4D radiation imaging.
키워드
- 제목
- CAD 모델 기반의 4D CT 데이터 제작 의용공학 융합 프로그램 개발
- 제목 (타언어)
- Development of 4D CT Data Generation Program based on CAD Models through the Convergence of Biomedical Engineering
- 저자
- 서정민; 한민철; 이현수; 이세형; 김찬형
- 발행일
- 2017-04
- 저널명
- 한국융합학회논문지
- 권
- 8
- 호
- 4
- 페이지
- 131 ~ 137