Transmission efficiency of the recoil mass spectrometer EMMA at TRIUMF

  • Davids, B.; 
  • Esker, N. E.; 
  • Jaeyoung, J.; 
  • Kim, Y. K.; 
  • Pak, K.; 
  • 외 1명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

The mean transmission efficiency of the EMMA recoil mass spectrometer at TRIUMF has been measured with 6 different angular apertures at 17 kinetic energy/charge deviations with respect to the central, reference trajectory. Measurements performed using a 148Gd α source installed at the target position of the spectrometer are compared to ion-optical calculations and Monte Carlo simulations. The transmission efficiency as a function of angle and kinetic energy/charge is described empirically using piecewise Gaussian functions whose parameters are fit to the data. © 2026

키워드

Recoil mass spectrometer; Transmission efficiency; ISAC-II
제목
Transmission efficiency of the recoil mass spectrometer EMMA at TRIUMF
저자
Davids, B.; Esker, N. E.; Jaeyoung, J.; Kim, Y. K.; Pak, K.; Williams, M.
DOI
10.1016/j.nima.2026.171295
발행일
2026-05
유형
Article
저널명
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
권
1085
페이지
1 ~ 11