Cerebrospinal fluid-compensated medication reservoir for an implantable infusion device: Concept and preliminary evaluation

  • Nam, Kyoung Won
  • Choi, Seong Wook
  • Kim, In Young
  • Kim, Kwang Gi
  • Jo, Yung Ho
  • 외 1명
Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

Conventional gas-compensated medication reservoirs used for implantable infusion devices require perfect sealing of the gas chamber, because the gases used are generally toxic. In addition, the physical properties of selected gas critically affect the performance of infusion devices and hydraulic performance of the infusion device can be affected by the amount of medication discharged. In this study, we suggest a new medication reservoir that adopts a cerebrospinal fluid (CSF)-compensating mechanism, such that when a medication is released from the reservoir by a mechanical actuator, native CSF enters into the reservoir to minimize the build-up of pressure drop. We evaluated in vitro performance and conducted in vivo feasibility tests by using an intrathecal infusion device developed at the Korean National Cancer Center. Experimental results showed that the proposed CSF-compensated infusion pump was essentially less affected by ambient temperature or pressure conditions compared to the gas-compensated infusion pump. Moreover, it showed moderate implant feasibility and operating stability during an animal experiment performed for 12 days. We believe that the proposed volume-compensating mechanism could be applied in various medical fields that use implantable devices.

키워드

Cerebrospinal fluidImplantable infusion deviceIntrathecal infusion deviceMedication reservoirVolume compensationREFRACTORY CANCER PAININTRATHECAL DRUG-DELIVERYPUMPBRAINMANAGEMENTSURVIVALTOXICITY
제목
Cerebrospinal fluid-compensated medication reservoir for an implantable infusion device: Concept and preliminary evaluation
저자
Nam, Kyoung WonChoi, Seong WookKim, In YoungKim, Kwang GiJo, Yung HoKim, Dae Hyun
DOI
10.5301/ijao.5000138
발행일
2013-05
유형
Article
저널명
International Journal of Artificial Organs
36
5
페이지
350 ~ 357