상세 보기
Development of custom-made RF coil for magnetic resonance velocimeter with a high spatial resolution
- Yang, Byungkuen;
- Cho, Jee-Hyun;
- Lee, Jeesoo;
- Song, Simon
WEB OF SCIENCE
4SCOPUS
5초록
Magnetic resonance velocimetry (MRV) is a versatile flow visualization technique that is used to measure three-component velocity vectors in a 3D space. However, the spatial resolution of MRV is relatively poor in comparison with optical flow visualization techniques, thereby limiting its applicability to small-scale flows and wall shear stress (WSS) estimation. Thus, we built a solenoid RF coil and evaluated its performance in terms of spatial resolution improvement by measuring the laminar flow in a circular tubing. The coil was developed for a 4.7 Tesla MRI system and for tightly wrapping a flow tubing with an inner diameter of 2 mm. The custom-made RF coil could precisely capture the velocity vectors in a voxel that was 11 times smaller than a commercial coil at the same SNR. Therefore, the WSS errors estimated using the custom-made and commercial coils were 8.5 % and 42.3 %, respectively.
키워드
- 제목
- Development of custom-made RF coil for magnetic resonance velocimeter with a high spatial resolution
- 저자
- Yang, Byungkuen; Cho, Jee-Hyun; Lee, Jeesoo; Song, Simon
- 발행일
- 2019-04
- 유형
- Article
- 권
- 33
- 호
- 4
- 페이지
- 1681 ~ 1688