Accurate measurement of uranium and thorium in naturally occurring radioactive materials to overcome complex matrix interference

  • Lee, JuHyun
  • Yi, Sung-Chul
Citations

WEB OF SCIENCE

3
Citations

SCOPUS

4

초록

A standard addition (SA) calibration method for determining uranium and thorium in naturally-occurring radioactive materials (NORMs) was studied. Pretreatment method using fusion was established and optimized. The SA method was validated (LOD, LOQ, linearity, selectivity, and accuracy) using certified reference materials (SRM 2709a, RM-ZR, and RM-BX). The results were evaluated for bias, and a correction equation that included a contribution to the expanded uncertainty was used. All performance criteria were satisfied; the results agreed well with certified and reference values.

키워드

UraniumThoriumStandard additionICP-MSValidationNORMMEASUREMENT UNCERTAINTYICP-MSSTANDARD ADDITIONSMETHOD VALIDATIONSPECTROMETRYSAMPLESRADIONUCLIDESINFORMATIONPROFILESRECOVERY
제목
Accurate measurement of uranium and thorium in naturally occurring radioactive materials to overcome complex matrix interference
저자
Lee, JuHyunYi, Sung-Chul
DOI
10.1016/j.apradiso.2023.110649
발행일
2023-03
유형
Article
저널명
Applied Radiation and Isotopes
193
페이지
1 ~ 7