Anti-Reflux Ureteral Stent with Polymeric Flap Valve Using Three-Dimensional Printing: An In Vitro Study

  • Park, Chang-Ju
  • Kim, Hyeon-Woo
  • Jeong, Sangdo
  • Seo, Seungwan
  • Park, Yangkyu
  • ... Moon, Hong Sang
  • 외 1명
Citations

WEB OF SCIENCE

34
Citations

SCOPUS

41

초록

Purpose: This article aims to describe the design of an anti-reflux ureteral stent with a polymeric flap valve and the fabrication methods using three-dimensional (3D) printing. The stent effectively prevents backward flow with a negligible reduction in forward flow. Fabrication of miniaturized valves was easy with high precision and rapid prototyping. Materials and Methods: The proposed stent comprised a 7F Double-J (DJ) stent and a polymeric flap valve. The valve was made of Tango Plus FLX980 and was fabricated using a 3D printer. Two types of stent were prepared for in vitro tests: DJ stents with (1) an uncoated valve (UCV) stent and (2) a parylene C coated valve (PCV) stent for enhanced biocompatibility. The flow characteristics of each stent were evaluated considering flow direction, parylene coating, and stent side holes, and were compared to the intact DJ stent. Results: The forward flow rate for the distal portion of the UCV and PCV stents was 9.8mL/min and 7.8mL/min at applied pressure of 15cm H2O (normal anterograde pressure in patients with stents), respectively. Backward flow rate for the distal portion of the UCV and PCV stents was decreased by 28 times and 8 times at applied pressure of 50cm H2O (maximum bladder pressure), respectively, compared with the distal portion of the intact DJ stent. Forward flow rates of whole stents were 22.2mL/min (UCV stent) and 20.0mL/min (PCV stent) at applied pressure of 15cm H2O, and backward flow rates of whole UCV and PCV stents were decreased by 8.3 times and 4.0 times at applied pressure of 50cm H2O, respectively, compared with the intact DJ stent. Conclusions: The anti-reflux ureteral stent was successfully designed and fabricated using a 3D printer. In vitro studies showed that the stent effectively prevented backward flow while minimizing reduction in forward flow.

키워드

EXPERIENCECOMPLICATIONSSYMPTOMSFLOW
제목
Anti-Reflux Ureteral Stent with Polymeric Flap Valve Using Three-Dimensional Printing: An In Vitro Study
저자
Park, Chang-JuKim, Hyeon-WooJeong, SangdoSeo, SeungwanPark, YangkyuMoon, Hong SangLee, Jong-Hyun
DOI
10.1089/end.2015.0154
발행일
2015-08
유형
Article
저널명
Journal of Endourology
29
8
페이지
933 ~ 938