Blood compatibility of novel zwitterionic PEG-grafted nitinol alloy for peripheral arterial stents

Citations

SCOPUS

4

초록

In this study, newly synthesized zwitterionic PEG was grafted on Nitinol alloy using oxidation treatment. The surface property and blood compatibility of surface-modified Nitinols were examined. The results of surface analysis showed that the contact angle and the ratio of oxygen to carbon significantly decreased with Nitinol alloys. The total amount of fibrinogen (0.095 μg/cm2) adsorbed onto TiNi-PEG2K-N+-S- was lower than that of TiNi control (0.12 μg/cm2). The platelet adhesion decreased in the order of TiNi control > TiNi-MPEG2K > TiNi-PEG2K-N+-S. Particularly, zwitterionic PEG with PEG2K was proven better than any others. The results indicated that zwitterionic PEG surface could significantly suppress platelet adhesion and protein adsorption as compared to other samples. The present study suggested that grafted zwitterionic PEG structure may possess improved blood compatibility.

키워드

Blood compatibilityNitinol alloyStentSurface modificationZwitterionic PEGBiocompatibilityBiosynthesisOxidationPolyethylene glycolsSurface propertiesSurface treatmentBlood compatibilityNitinol alloysOxidation treatmentZwitterionic PEGBiomedical engineering
제목
Blood compatibility of novel zwitterionic PEG-grafted nitinol alloy for peripheral arterial stents
저자
Kim, Jae HoonSon, Jun SikPark, KwideokKim, Jong ManKim, Jae-JinAhn, Kwang-DukHan, Dong Keun
DOI
10.4028/0-87849-436-7.805
발행일
2007-09
유형
Conference Paper
저널명
Key Engineering Materials
342-343
페이지
805 ~ 808