Feasibility Study on Active Back Telemetry and Power Transmission Through an Inductive Link for Millimeter-Sized Biomedical Implants

Citations

WEB OF SCIENCE

49
Citations

SCOPUS

48

초록

This paper presents a feasibility study of wireless power and data transmission through an inductive link to a 1-mm(2) implant, to be used as a free-floating neural probe, distributed across a brain area of interest. The proposed structure utilizes a four-coil inductive link for back telemetry, shared with a three-coil link for wireless power transmission. We propose a design procedure for geometrical optimization of the inductive link in terms of power transmission efficiency (PTE) considering specific absorption rate and data rate. We have designed a low-power pulse-based active data transmission circuit and characterized performance of the proposed inductive link in terms of its data rate and bit error rate (BER). The 1-mm(2) data-Tx/power-Rx coil is implemented using insulated bonding wire with 25-mu m diameter, resulting in measured PTE in tissue media of 2.01% at 131 MHz and 1.8-cm coil separation distance when the resonator coil inner radius is 1 cm. The measured BER at 1-Mbps data rate was 2.7 x 10(-6) and 5 x 10(-6) in the air and tissue environments, respectively.

키워드

Free-floating neural interfacing probes; mm-sized implantable medical devices; pulse-based near-field transceiver; wireless power/data transmission; WIRELESS; TRANSCEIVER; INTERFACES; DESIGN; DEVICE
제목
Feasibility Study on Active Back Telemetry and Power Transmission Through an Inductive Link for Millimeter-Sized Biomedical Implants
저자
Yeon, Pyungwoo; Mirbozorgi, S. Abdollah; Lim, Jaemyung; Ghovanloo, Maysam
DOI
10.1109/TBCAS.2017.2775638
발행일
2017-12
유형
정기학술지(Article(Perspective Article포함))
저널명
IEEE Transactions on Biomedical Circuits and Systems
권
11
호
6
페이지
1366 ~ 1376