저간섭성반사계와 부분결합간섭계를 이용한 안구생체계측과 백내장 수술 후 굴절력 예측의 비교

Comparison of Ocular Biometry and Postoperative Refraction in Cataract Patients Between Lenstar® and IOL Master®
  • 신중원
  • 성민철
  • 강민호
  • 조희윤
  • 이윤정

초록

Purpose: To compare axial length, anterior chamber depth, and keratometric measurements of an optical low-coherence reflectometry device with those of other ocular biometry devices and evaluate the accuracy of predicting postoperative refraction. Methods: A total of 32 eyes in 32 patients who received cataract surgery were included in the present study. The axial length, anterior chamber depth, and keratometry were measured by optical low-coherence reflectometry (Lenstar LS900®), partial coherence interferometry (IOL master®), and ultrasound. The SRK/T formula was used to calculate IOL power, and predictive error that subtracts predictive refraction from postoperative refraction was compared among ocular biometry devices. Results: Axial length, anterior chamber depth, and keratometry had a strong correlation and demonstrated no statistically significant differences between Lenstar LS900® and other devices. The Bland–Altman plots showed a high degree of agreement between Lenstar LS900® and other devices. The mean absolute prediction errors in Lenstar LS900® and IOL master® were not significantly different. Conclusions: The ocular biometric measurements and prediction of postoperative refraction using Lenstar LS900® were as accurate as IOL master® and ultrasound.

키워드

Lenstar LS900Ocular biometryOptical low coherence reflectometry
제목
저간섭성반사계와 부분결합간섭계를 이용한 안구생체계측과 백내장 수술 후 굴절력 예측의 비교
제목 (타언어)
Comparison of Ocular Biometry and Postoperative Refraction in Cataract Patients Between Lenstar® and IOL Master®
저자
신중원성민철강민호조희윤이윤정
DOI
10.3341/jkos.2012.53.6.833
발행일
2012-06
저널명
대한안과학회지
53
6
페이지
833 ~ 838