상세 보기
Rapid detection of Mycoplasma pneumonia in a microfluidic device using immunoagglutination assay and static light scattering
- Kim, Keesung;
- Jung, Ho-Sup;
- Song, Jae-Young;
- Lee, Man-Ryul;
- Kim, Kye-Seong;
- 외 1명
Citations
WEB OF SCIENCE
10Citations
SCOPUS
10초록
We present real-time, rapid detection of Mycoplasma pneumonia in PBS inside a Y-channel PDMS microfluidic device via optical fiber monitoring of latex immunoagglutination. The latex immunoagglutination assay was performed with serially diluted M. pneumonia solutions using highly carboxylated polystyrene particles of 390 and 500 nm diameter conjugated with monoclonal anti-M. pneumonia. Proximity optical fibers were located around the viewing cell of the device, which were used to measure the increase in 45 degrees forward light scattering of the aggregated particles. The detection limit are slightly less than 50 pg m L-1 both for 390 and 500 nm microspheres and the detection time do not exceed 90 s.
키워드
Cell culture; Fiber optic sensor; Immunoagglutination assay; Microfluidic device; Mycoplasma pneumonia; SOFT-LITHOGRAPHY; CELL-CULTURES; CONTAMINATION; MICROSPHERES; SYSTEM; SIZE
- 제목
- Rapid detection of Mycoplasma pneumonia in a microfluidic device using immunoagglutination assay and static light scattering
- 저자
- Kim, Keesung; Jung, Ho-Sup; Song, Jae-Young; Lee, Man-Ryul; Kim, Kye-Seong; Suh, Kahp-Yang
- 발행일
- 2009-09
- 유형
- Article
- 저널명
- Electrophoresis
- 권
- 30
- 호
- 18
- 페이지
- 3206 ~ 3211