Establishmet of Reproducible Ultra-High-Dose Rate Proton Beams Based on a Clinical Cyclotron System

  • Kim Chae-Eon
  • Park, Sang-il
  • Jeong, Seonghoon
  • An, Seohyeon
  • Kim, Chankyu
  • ... Chung, Yoonsun
  • 외 5명

초록

Purpose: Stable delivery of ultra-high-dose rates (UHDRs) proton beams for FLASH research remains technically challenging. This study aimed to develop a UHDR proton beam on a clinical cyclotron with minimal beamline modifications and a time-gated control approach for experimental and preclinical FLASH research. Methods: UHDR proton beams were generated using a cyclotron-driven passive-scattering system without the range modulator and scatterer. Field broadening was achieved with thin lead foil, while a Raspberry Pi-based time-gated system was implemented to control beam-on duration independent of the monitor chamber. Beam profiles were evaluated using EBT-XD films. Results: Millisecond-scale beam delivery with high temporal reproducibility was attained for beam-on durations. A stable reference UHDR beam was established at 228 MeV and 200 nA, producing a Gaussian-like lateral profile with a 28.8 mm full width at half maximum at 2 g/cm2. Film dosimetry confirmed a central-axis dosage rate of approximately 70 Gy/s. Conclusions: This study demonstrates that clinically available clinical cyclotron systems can be adapted to generate UHDR proton beams suitable for FLASH research through time-gated control and minimal beamline modifications. The established reference beam configuration has potential for experimental and preclinical FLASH proton studies.

키워드

Ultra-high dose ratesFLASH proton therapyCyclotron-based proton therapy system
제목
Establishmet of Reproducible Ultra-High-Dose Rate Proton Beams Based on a Clinical Cyclotron System
저자
Kim Chae-EonPark, Sang-ilJeong, SeonghoonAn, SeohyeonKim, ChankyuJeong, Jong HwiKim, HaksooLim, Young KyungShin, DonghoChung, YoonsunLee, Se Byeong
DOI
10.14316/pmp.2026.37.2.54
발행일
2026-06
유형
Y
저널명
PROGRESS in MEDICAL PHYSICS
37
2
페이지
54 ~ 64