A Behind-The-Ear Patch-Type Mental Healthcare Integrated Interface with Adaptive Multimodal Offset Compensation and Parasitic Cancellation

  • Kim, Hyunjoong; 
  • Cho, Sanghyeon; 
  • Kim, Myeong Woo; 
  • Park, Chan Sam; 
  • Lee, Kwangmuk; 
  • ... Jang, Dong Pyo; 
  • 외 5명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

2

초록

A behind-the-ear (BTE) integrated interface for mental healthcare applications is presented, featuring optimized BTE electrode configurations and wide multimodal biomedical IC with adaptive compensation capabilities. The proposed IC supports 8 bio-potential (ExG), 1 photoplethysmogram (PPG), 1 galvanic skin response (GSR), 1 bio-impedance (BioZ), and 2 stimulation channels. The ExG channel achieves 2.5GΩ input impedance, boosted by 308 times with offset compensated auxiliary path (OCAP) architecture, and its AC input impedancecharacteristic is boosted further by dual resolution external positive feedback loop (DR-EPFL) scheme. An area and energy-efficient GSR-embedded ECG recording scheme is presented. For comprehensive multimodal sensing features, dual-slope PPG channel with parasitic capacitance compensation, electrode-tissue impedance adaptive stimulator, and high dynamic range BioZ channel are integrated. The IC was fabricated in a 0.18-μm BCD process and integrated into a BTE patch-type device prototype. System-level feasibility was experimentally verified through in-vivo stress measurements with virtual reality (VR) environment, demonstrating effective mental health monitoring capabilities.

키워드

Electrodes; Recording; Electrocardiography; Electroencephalography; Monitoring; Impedance; Medical services; Biomedical measurement; Biomedical monitoring; Printed circuits; Behind-the-ear (BTE); multimodal biomedical system; readout integrated circuit (ROIC); electrode artifact; EEG; ECG; PPG; GSR; taVNS; stress measurement; TO-DIGITAL CONVERTER; STIMULATION ARTIFACT; CHOPPER AMPLIFIER; POWER MANAGEMENT; INPUT IMPEDANCE; RANGE; SOC
제목
A Behind-The-Ear Patch-Type Mental Healthcare Integrated Interface with Adaptive Multimodal Offset Compensation and Parasitic Cancellation
저자
Kim, Hyunjoong; Cho, Sanghyeon; Kim, Myeong Woo; Park, Chan Sam; Lee, Kwangmuk; Song, Solwoong; Keum, Dae Sik; Lee, Sangmoon; Jeong, Hoon Eui; Jang, Dong Pyo; Kim, Jae Joon
DOI
10.1109/TBCAS.2025.3642345
발행일
2026-04
유형
Article
저널명
IEEE Transactions on Biomedical Circuits and Systems
권
20
호
2
페이지
313 ~ 327