Prototype system for proton beam range measurement based on gamma electron vertex imaging

  • Lee, Han Rim
  • Kim, Sung Hun
  • Park, Jong Hoon
  • Jung, Won Gyun
  • Lim, Hansang
  • ... Kim, Chan Hyeong
Citations

WEB OF SCIENCE

15
Citations

SCOPUS

15

초록

In proton therapy, for both therapeutic effectiveness and patient safety, it is very important to accurately measure the proton dose distribution, especially the range of the proton beam. For this purpose, recently we proposed a new imaging method named gamma electron vertex imaging (GEVI), in which the prompt gammas emitting from the nuclear reactions of the proton beam in the patient are converted to electrons, and then the converted electrons are tracked to determine the vertices of the prompt gammas, thereby producing a 2D image of the vertices. In the present study, we developed a prototype GEVI system, including dedicated signal processing and data acquisition systems, which consists of a beryllium plate (= electron converter) to convert the prompt gammas to electrons, two double-sided silicon strip detectors (= hodoscopes) to determine the trajectories of those converted electrons, and a plastic scintillation detector (= calorimeter) to measure their kinetic energies. The system uses triple coincidence logic and multiple energy windows to select only the events from prompt gammas. The detectors of the prototype GEVI system were evaluated for electronic noise level, energy resolution, and time resolution. Finally, the imaging capability of the GEVI system was tested by imaging a Sr-90 beta source, a Co-60 gamma source, and a 45-MeV proton beam in a PMMA phantom. The overall results of the present study generally show that the prototype GEVI system can image the vertices of the prompt gammas produced by the proton nuclear interactions.

키워드

Proton therapyProton dose monitoringBeam range measurementPrompt gammaGamma electron vertex imagingENERGY CALIBRATIONCOMPTON CAMERASLIT CAMERAVERIFICATION
제목
Prototype system for proton beam range measurement based on gamma electron vertex imaging
저자
Lee, Han RimKim, Sung HunPark, Jong HoonJung, Won GyunLim, HansangKim, Chan Hyeong
DOI
10.1016/j.nima.2017.03.022
발행일
2017-06
유형
Article
저널명
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
857
페이지
82 ~ 97