Enhancement of Collection Efficiency of an Inertial Impactor Using an Additional Punched Impaction Plate

  • Cheon, Tae-Won
  • Lee, Jong-Yeol
  • Bae, Jae-Young
  • Yook, Se-Jin
Citations

WEB OF SCIENCE

13
Citations

SCOPUS

13

초록

In this study, the collection efficiency of a round-nozzle inertial impactor was enhanced through use of an additional punched impaction plate. The additional plate was installed between the nozzle outlet and the existing impaction plate. To determine its effect, a simulation and an experiment were conducted. A parametric study was performed on a PM2.5 inertial impactor to improve the optimum efficiency based on the hole diameter of the additional plate and its distance from the existing plate. As a result, the cut-off size was reduced by 27% from 2.52 mu m to 1.84 mu m. The effect of the additional punched impaction plate on PM1 and PM10 inertial impactors was also investigated. The improvement in efficiency could be observed even when the additional punched impaction plate had varying Reynolds numbers and nozzle sizes. Accordingly, the (Stk(50))(0.5) value was reduced by 25% from 0.49 to 0.37 by adding a punched impaction plate in a roundnozzle inertial impactor.

키워드

Inertial impactorCollection efficiencyStokes numberAdditional punched impaction plateROUND-NOZZLE IMPACTORSPARTICLESPERFORMANCEINLET
제목
Enhancement of Collection Efficiency of an Inertial Impactor Using an Additional Punched Impaction Plate
저자
Cheon, Tae-WonLee, Jong-YeolBae, Jae-YoungYook, Se-Jin
DOI
10.4209/aaqr.2017.01.0018
발행일
2017-10
유형
Article
저널명
Aerosol and Air Quality Research
17
10
페이지
2349 ~ 2357