Fixed volume sequential standard addition calibration: Value assignment of impurities in zero gas

Citations

WEB OF SCIENCE

4
Citations

SCOPUS

4

초록

A fixed volume case of sequential standard addition calibration (S-SAC) for the value assignment of impurities in zero gas is described. A mathematical description of this technique has been derived and has been shown to exhibit a similar systematic bias during the extrapolation process to that seen for other S-SAC cases. The use of S-SAC in the gas phase has demonstrated that variation in the sample volume for S-SAC in the liquid phase is not the generalised cause of the bias experienced during extrapolation. Instead it is the variation in the volume of the original sample with respect to the overall volume of the mixture following the addition of standard. In addition, the requirement to perform a bias correction on the extrapolated values has been discussed and best practice solutions for correction proposed.

키워드

standard additioncalibrationbiaszero gasmatrixSYSTEMATIC-ERROR
제목
Fixed volume sequential standard addition calibration: Value assignment of impurities in zero gas
저자
Brown, Richard J. C.Brewer, Paul J.Kim, Ki-Hyun
DOI
10.1016/j.chemolab.2017.03.012
발행일
2017-05
유형
Article
저널명
Chemometrics and Intelligent Laboratory Systems
164
페이지
52 ~ 55