Monitoring for telescope array fluorescence detector PMT camera by YAP and Xe flasher
- Shin, Bokkyun;
- Tokuno, H.;
- Ikeda, Daisuke;
- Fukushima, Masaki;
- Tsunesada, Yoshiki;
- ... Cheon, Byung Gu;
- 외 1명
SCOPUS
0초록
Gain stability of the photomultiplier (PMT) cameras for The Fluorescence Detectors (FD) in the Telescope Array experiment has been monitored by using Am-241 loaded scintillator pulsers (YAP) and diffused xenon flashers (TXF). Each FD camera consists of 16 × 16 PMT(Hamamatsu-R9508) array, and contains two or three absolutely calibrated PMTs carrying a YAP. YAP, which is attached on an PMT cathode, is a small light source which emits UV photons at 50Hz. A TXF which is a relative calibration installed on each camera shoots photons simultaneously on all the 256 PMTs uniformly from about 3 meter away. From the monitoring of YAP pulsers over 4 years of FD operation we found an annual average of slow monotonic PMT gains drift is 0.01 ± 0.31(stat.) ± 0.21(syst.)% and it’s sigma is 1.7%. In conclusion, no systematic PMT gain decrease by the night sky background was observed. A gains monitoring by the TXF also shows the same PMT gain drifts within the annual magnitude of 0.86 ± 0.13(stat.)%/yearl.