Development of an array-type prompt gamma detection system for the online measurement of the range of the proton beam in a patient: a Monte Carlo feasibility study

Citations

WEB OF SCIENCE

15
Citations

SCOPUS

15

초록

Recently a prototype prompt gamma scanner system, named the PGS (prompt gamma scanner), was constructed and used to demonstrate that there is a clear correlation between prompt gamma distribution and the range of the proton beam in a patient. However, the PGS system requires a scanning process to obtain the prompt gamma distribution; consequently, it cannot be used in "spot scanning", in which the proton beam is continuously moving during treatment. The present Monte Carlo study explored the feasibility of a new detection system that does not require a scanning process, that is, that uses a linear array of scintillation detectors and photodiodes for online measurement of the proton beam range. The simulation results are very encouraging, showing that an array-type prompt gamma detection system can measure the proton beam range despite the fact that, with this system, the level of background gammas somewhat increases as a result of reduced collimator shielding. This paper discusses those simulation results, as well as some important design considerations based on them.

키워드

proton therapybeam rangeprompt gammasMonte Carloradiation dosimetryPET
제목
Development of an array-type prompt gamma detection system for the online measurement of the range of the proton beam in a patient: a Monte Carlo feasibility study
저자
Min, Chul HeePark, Jang GuenKim, Chan Hyeong
DOI
10.3938/jkps.52.888
발행일
2008-03
유형
Article
저널명
Journal of the Korean Physical Society
52
3
페이지
888 ~ 891