상세 보기
Development of an array-type prompt gamma detection system for the online measurement of the range of the proton beam in a patient: a Monte Carlo feasibility study
- Min, Chul Hee;
- Park, Jang Guen;
- Kim, Chan Hyeong
WEB OF SCIENCE
15SCOPUS
15초록
Recently a prototype prompt gamma scanner system, named the PGS (prompt gamma scanner), was constructed and used to demonstrate that there is a clear correlation between prompt gamma distribution and the range of the proton beam in a patient. However, the PGS system requires a scanning process to obtain the prompt gamma distribution; consequently, it cannot be used in "spot scanning", in which the proton beam is continuously moving during treatment. The present Monte Carlo study explored the feasibility of a new detection system that does not require a scanning process, that is, that uses a linear array of scintillation detectors and photodiodes for online measurement of the proton beam range. The simulation results are very encouraging, showing that an array-type prompt gamma detection system can measure the proton beam range despite the fact that, with this system, the level of background gammas somewhat increases as a result of reduced collimator shielding. This paper discusses those simulation results, as well as some important design considerations based on them.
키워드
- 제목
- Development of an array-type prompt gamma detection system for the online measurement of the range of the proton beam in a patient: a Monte Carlo feasibility study
- 저자
- Min, Chul Hee; Park, Jang Guen; Kim, Chan Hyeong
- 발행일
- 2008-03
- 유형
- Article
- 권
- 52
- 호
- 3
- 페이지
- 888 ~ 891