Single-Shot Near-Field Volumetric Imaging System for Optical Ultrasound and Photoacoustics Using Capacitive Micromachined Ultrasonic Transducer Without Transmission Mode

  • Choi, Won Young
  • Kwon, Soo Won
  • Kim, Young Hun
  • Kang, Ki Chang
  • Park, Kwan Kyu
Citations

WEB OF SCIENCE

7
Citations

SCOPUS

8

초록

In this article, we present a single-shot dual-mode imaging system that uses optical ultrasound (US) as an ultrasonic pulser without a transmission circuit. The ultrasonic pulse-echo system comprises an optical US pulser generated by carbon nanotubes (CNTs), which generate a high-power photoacoustic (PA) signal and a capacitive micromachined ultrasonic transducer (CMUT) receiver. By fabricating a thin CNT-polydimethylsiloxane (PDMS) composite capable of semiabsorption of the laser, a single-shot imaging system was developed. By transmitting a semipenetration light to the object, US and PA imaging were performed in a single shot. A CNT thickness of <1 mu m produced a maximum pressure of 154 kPa, and US was received by CMUT with a 2-MHz center frequency in PDMS. Additionally, a low-profile and near-depth imaging system was constructed with an intermediate layer of the 6-mm PDMS for the dry contact method. We performed a single-shotdual-mode imaging experimenton point and line phantoms, as well as the particle spread in the soft tissue. Thus, we examined the feasibility of the near-depth and single-shot dual-mode (US and PA) imaging system capable of a dry contact.

키워드

Dual-mode imagingnear-field imagingphotoacoustic (PA)single shotultrasound (US) imagingHIGH-RESOLUTIONMICROSCOPYGENERATION
제목
Single-Shot Near-Field Volumetric Imaging System for Optical Ultrasound and Photoacoustics Using Capacitive Micromachined Ultrasonic Transducer Without Transmission Mode
저자
Choi, Won YoungKwon, Soo WonKim, Young HunKang, Ki ChangPark, Kwan Kyu
DOI
10.1109/TUFFC.2020.2965600
발행일
2020-06
유형
Article
저널명
IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
67
6
페이지
1151 ~ 1158