Repeatability of Scheimpflug-Placido camera in mild dry eye versus normal eyes according to the topographical position of the cornea

Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

We aimed to investigate the repeatability of various corneal measurements according to topographical location in the entire cornea measured by dual rotating Scheimpflug-Placido camera and to explore the differences in repeatability between patients with mild dry eye and those with normal eyes. In both the normal and dry eye groups, divided based on BUT or the height of the tear film, there were no statistically significant differences in the ratio of unacceptable variation (RUV) and ICC. The consistency of the examination of the anterior and posterior refractive values and corneal thickness according to the corneal location, measured three times repeatedly using the Galilei anterior segment camera, was high. There was no difference based on the height of the tear film or the tear film break-up time. However, caution is needed when interpreting the values of the anterior corneal refractive values, as there can be changes of more than 0.5D within 3 mm of the central area.

키워드

OPTICAL COHERENCE TOMOGRAPHYORBSCAN-IITHICKNESS MEASUREMENTREFRACTIVE SURGERYCLASSIFICATIONPOWERDEFINITIONPENTACAMGALILEISHAPE
제목
Repeatability of Scheimpflug-Placido camera in mild dry eye versus normal eyes according to the topographical position of the cornea
저자
Hwang, SunjinKim, Dae SungKim, DurooHong, Eun HeeShin, Yong UnKim, Yu JeongKang, Min Ho
DOI
10.1038/s41598-024-73900-z
발행일
2024-10
유형
Article
저널명
Scientific Reports
14
1
페이지
1 ~ 8