상세 보기
Effect of horizontal inlet on slit-nozzle virtual impactor performance
- Heo, Ji-Eun;
- Zahir, Muhammad Zeeshan;
- Park, Hyunjin;
- Seo, Jeonghyun;
- Park, Hyeongjun;
- ... Yook, Se-Jin
WEB OF SCIENCE
7SCOPUS
9초록
We observed variations in collection efficiency and wall loss in a virtual impactor by applying a horizontal inlet. A parametric study on the dimensions of the horizontal inlet was conducted to determine the shape that could maximize the collection efficiency and minimize the wall loss. As a conventional vertical inlet was replaced by a horizontal inlet, the PM2.5 virtual impactor decreased the cut-off size by 17 % from 2.5 μm to 2.08 μm while the PM10 virtual impactor decreased the cut-off size by 22 % from 10 μm to 7.8 μm. The application of a horizontal inlet significantly reduced the wall loss for a particle size corresponding to a particular cut-off size, overcoming the problem with the existing virtual impactor.
키워드
- 제목
- Effect of horizontal inlet on slit-nozzle virtual impactor performance
- 저자
- Heo, Ji-Eun; Zahir, Muhammad Zeeshan; Park, Hyunjin; Seo, Jeonghyun; Park, Hyeongjun; Yook, Se-Jin
- 발행일
- 2018-05
- 유형
- Article
- 권
- 32
- 호
- 5
- 페이지
- 2419 ~ 2424