Development of a beta gauge system for a fabric density measurement

  • Kim, Han Soo
  • Park, Se Hwan
  • Ha, Jang Ho
  • Song, Tae Young
  • Cho, Seung Yeon
  • ... Kim, Yong Kyun
Citations

WEB OF SCIENCE

7
Citations

SCOPUS

9

초록

A beta gauge system was designed and fabricated for a fabric density measurement with high sensitivity in real time. An ionization chamber and 3.7 GBq Kr-85 was chosen as a beta radiation sensor and a beta source. An ionization chamber, which had a 0.5 mu m-thick aluminized mylar film as an incident window, was constructed with a guard electrode structure in consideration of minimizing leakage currents. Xe gas was chosen as a filling gas to increase sensitivity of an ionization chamber. Linearity of the ionization currents against the densities was 0.9894 in the selected measurement range. And the deviation of the ionization currents at a standard density was below 0.02%. The ranges of the density measurement and the active area were from 20 to 999 g/cm(2) with +/- 1% accuracy and 40 mm diameters. The fabricated beta gauge system can be applied to fields such as in the paper, film, and plastic production industries.

키워드

Beta gaugeKr-85Ionization chamberDensity measurementNDTDensitometersDensity measurement (optical)Density measurement (specific gravity)FabricsGagesIonization chambersKryptonLeakage currentsXenonkrypton 85xenonKrActive areaAluminized MylarBeta gaugeBeta gaugesBeta radiationDensity measurementDensity MeasurementsFilling gasGuard electrodesHigh sensitivityIonization currentMeasurement rangeNDTPlastic productionReal timearticlebeta radiationdensitometryelectric currentequipment designionization chamberpriority journalradiation monitorradiation monitoringtextileIonization of gases
제목
Development of a beta gauge system for a fabric density measurement
저자
Kim, Han SooPark, Se HwanHa, Jang HoSong, Tae YoungCho, Seung YeonKim, Yong Kyun
DOI
10.1016/j.apradiso.2009.02.015
발행일
2009-07
유형
Article; Proceedings Paper
저널명
Applied Radiation and Isotopes
67
7-8
페이지
1213 ~ 1215