위전도 측정을 위한 전극간 부착거리에 관한 연구

Influences of Inter-electrode Distance on Electrogastrography Measurements

초록

Cutaneous electrogastrography is the measurement of electrical activity of the stomach on the abdominal surface. The validity of cutaneous electrogastrography is dependent upon the quality of the recording technique. The locations of electrodes are an important issue. We examined the influences of the inter-electrode distance of bipolar leads on electrogastrography measurements. The sensitivity distributions of EGG leads were calculated based on a 2D body fat model and evaluated according to the region of interest sensitivity ratio (ROISR). We simulated the ROISR of the inter-electrode distance in relation to various body fat thicknesses. The distance between the electrodes was proportional to the distance between the ROI and the surface of the abdomen. The results imply that inter-electrode distance can be applied in electrogastrography according to human body fat thickness.

키워드

ElectrogastrographySurface electrodesbipolar recordingsurface potentialfat thicknessElectrogastrographySurface electrodesbipolar recordingsurface potentialfat thickness
제목
위전도 측정을 위한 전극간 부착거리에 관한 연구
제목 (타언어)
Influences of Inter-electrode Distance on Electrogastrography Measurements
저자
한완택송인호김인영
발행일
2009-08
저널명
의공학회지
30
4
페이지
341 ~ 346