Development of light and radiation fields coincidence test toolkit for LINAC quality assurance

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

The coincidence of light and radiation fields is an important quality assurance test for accurate beam delivery in a linear accelerator (LINAC), and is recommended to be checked monthly by the American Association of Physicists in Medicine (AAPM). The test-related conventional tools may yield different results, especially without a commercial software, depending on the user, even if the tolerance suggested by the AAPM is very tight. This study was conducted to develop a semiautomatic film analyzer, called FSChecker, for a simple, easy, and consistent quantitative evaluation of the LINAC light and radiation field coincidence, without commercial tools. The FSChecker has two components, a dedicated grid paper and an analyzing software based on MATLAB. Comparing the results obtained from the FSChecker with those of a commercial software revealed that the FSChecker could provide a reasonable and reliable analysis for the coincidence test of light and radiation fields in a LINAC with various scan parameters, including an office-level configuration.

키워드

Linear acceleratorQuality assuranceLight fieldRadiation fieldLINEAR ACCELERATORSONCOLOGYFILM
제목
Development of light and radiation fields coincidence test toolkit for LINAC quality assurance
저자
Lee, Se HyungCheon, Bo-WiMin, Chul HeeKim, Chan HyeongKim, SeonghoonHan, Min Cheol
DOI
10.1007/s40042-022-00528-5
발행일
2022-08
유형
Article; Early Access
저널명
Journal of the Korean Physical Society
81
4
페이지
361 ~ 367