Semi-automated diagnostic system and severe fever with thrombocytopenia syndrome (SFTS) viral RNA reference materials encapsulated in virus-like particles for on-site diagnosis of SFTS

  • Roh, Yeonjeong
  • Baek, Yujin
  • Kim, Myoung Gyu
  • Koo, Bonhan
  • Seo, Ramin
  • ... Lee, Joonseok
  • 외 6명
Citations

WEB OF SCIENCE

7
Citations

SCOPUS

6

초록

Severe Fever with Thrombocytopenia Syndrome (SFTS), caused by the SFTS virus (SFTSV), is a tick-borne infectious disease prevalent in East Asia, characterized by acute fever and significant thrombocytopenia. Early and accurate diagnosis of SFTS is critical, particularly in rural areas where laboratory facilities are limited. Current diagnostic approaches, including RT-qPCR and immunological tests, face challenges due to their complexity, low sensitivity and the need for laboratory infrastructure. To address this, we developed a semi-automated diagnostic system for on-site testing of SFTS, integrating nucleic acid extraction with rapid, naked-eye detection using a reverse transcription loop-mediated isothermal amplification coupled with lateral flow assay. This system uses innovative SFTSV RNA reference materials encapsulated in virus-like particles, ensuring both stability and homogeneity, making them ideal for calibrating and validating diagnostic assays. Additionally, the semi-automated system, equipped with a microfluidic chip and optimized extraction processes, demonstrates a limit of detection (LOD) of 1.82 x 102 copies/mu L with SFTSV RNA reference materials. Finally, this system allows for the rapid and accurate detection of SFTSV RNA from 12 clinical specimens within 100 minutes, with results visible to the naked eye in just 2 minutes post-amplification. This system offers a practical solution for on-site SFTS diagnosis, particularly in resource-limited settings, and represents a significant advancement in the field of molecular diagnostics.

키워드

Severe fever with thrombocytopenia syndromeReference materialsVirus-like particleOn-site testMolecular diagnosticsNaked-eye detectionDROPLET DIGITAL PCRLENTIVIRUS VECTORINTERNATIONAL STANDARDS1STQUANTIFICATIONOPTIMIZATIONBUNYAVIRUSESTABLISHQUALITYDNA
제목
Semi-automated diagnostic system and severe fever with thrombocytopenia syndrome (SFTS) viral RNA reference materials encapsulated in virus-like particles for on-site diagnosis of SFTS
저자
Roh, YeonjeongBaek, YujinKim, Myoung GyuKoo, BonhanSeo, RaminLee, JoonseokJung, SeungwonKim, Sang KyungBae, MoonsukLee, Chang-SeopKim, Il-HwanShin, Yong
DOI
10.1016/j.snb.2025.137530
발행일
2025-06
유형
Article
저널명
Sensors and Actuators, B: Chemical
433
페이지
1 ~ 16