상세 보기
초록
When wireless devises for MICS(Medical Implant Communication Service) or ISM(Industrial Scientific and Medical) bands are designed, it is necessary to verify the performance by using a human body flat phantom. However, most of studies on the phantom are limited to the biological effects of mobile-phone EMF. In this paper, semi-solid phantoms having the electric properties suggested by FCC at MICS and ISM bands are fabricated. The manufactured phantoms satisfy the electric properties(εr=56.7 and σ=0.94 at MICS band, εr=52.7 and σ=1.95 at ISM band) at each band. All the composing materials for phantoms are commercially available in domestic market. Two methods using both polyethylene powder and TX-151 and glycerin at each band are proposed for diverse purpose. The electrical properties of the fabricated phantoms are measured by a dielectric probe kit and network analyzer after the lapse of one day (24 hours).
키워드
- 제목
- MICS 대역과 ISM 대역에서 인체 전기적 상수를 갖는 준(準) 고체형 플랫 팬텀 제작
- 제목 (타언어)
- The Study on Implementation of a Semi-Solid Flat Phantom with Equivalent Electrical Properties to Whole Human Body at MICS and ISM Band
- 저자
- 이순용; 서원범; 권결; 최재훈
- 발행일
- 2012-01
- 저널명
- 한국전자파학회 논문지
- 권
- 23
- 호
- 1
- 페이지
- 101 ~ 107