Compton imaging is a promising imaging modality for radioactive contamination in nuclear facilities. However, most of the Compton cameras in nuclear applications have been constructed with small detector sensors, up to a few centimeters, resulting in low imaging sensitivities. To overcome the limitation and improve imaging sensitivity, in this paper, a Compton camera, named a large-area Compton camera (LACC), was developed using two large-size (27 cm x 27 cm) monolithic Nal(TI) scintillators and 72 square-type photomultiplier tubes. The imaging performance of the LACC was evaluated for surface and internal contamination cases. Our results show that the absolute imaging sensitivity of the LACC was 3.4 x 10(-5) for a Cs-137 point source at about 1-m distance in front of the LACC, which is a few 10x higher than those of the existing Compton cameras. The obtained imaging resolution was 14.1 degrees full-width at half-maximum (FWHM) when we use the backprojection algorithm and 5.6 degrees FWHM when we use the maximum-likelihood expectation-maximization algorithm. The LACC also showed the 3-D imaging capability, not only for the surface contamination cases hut also for the internal contamination cases within a depth of a few tens of centimeters.
키워드
- 제목
- Large-Area Compton Camera for High-Speed and 3-D Imaging
- 저자
- Kim, Young-su; Kim, Jae Hyeon; Lee, Junyoung; Kim, Chan Hyeong
- 발행일
- 2018-11
- 유형
- Article
- 권
- 65
- 호
- 11
- 페이지
- 2817 ~ 2822