상세 보기
Study of the current-voltage characteristics of a SiC radiation detector irradiated by Co-60 gamma-rays
- Kang, Sang Mook;
- Ha, Jang Ho ;
- Park, Se Hwan;
- Kim, HS;
- Chun, Sung Dae;
- ... Kim, Yong Kyun
Citations
WEB OF SCIENCE
16Citations
SCOPUS
15초록
The SiC semiconductor has recently emerged as an attractive material for an ionization radiation detection. A wide bandgap (3.03 eV) and high radiation damage resistance allow for the fabrication of detectors capable of operating at a high-temperature and in high radiation fields. The major aim of our study is to develop a robust detector which will be applied to harsh radiation environments. In this study, we fabricated the SiC radiation detectors and measured the current-voltage characteristics of SiC detectors irradiated by Co-60 gamma-ray source. The I-V curves showed a decrease of the leakage current with increasing dose rate of Co-60 gamma-ray in the 0-100V bias voltage range.
키워드
SiC radiation detector; radiation damage; current-voltage characteristics; gamma-ray irradiation; semiconductor detector
- 제목
- Study of the current-voltage characteristics of a SiC radiation detector irradiated by Co-60 gamma-rays
- 저자
- Kang, Sang Mook; Ha, Jang Ho ; Park, Se Hwan; Kim, HS; Chun, Sung Dae; Kim, Yong Kyun
- 발행일
- 2007-08
- 유형
- Article; Proceedings Paper
- 권
- 579
- 호
- 1
- 페이지
- 145 ~ 147