Study of the current-voltage characteristics of a SiC radiation detector irradiated by Co-60 gamma-rays

  • Kang, Sang Mook
  • Ha, Jang Ho
  • Park, Se Hwan
  • Kim, HS
  • Chun, Sung Dae
  • ... Kim, Yong Kyun
Citations

WEB OF SCIENCE

16
Citations

SCOPUS

15

초록

The SiC semiconductor has recently emerged as an attractive material for an ionization radiation detection. A wide bandgap (3.03 eV) and high radiation damage resistance allow for the fabrication of detectors capable of operating at a high-temperature and in high radiation fields. The major aim of our study is to develop a robust detector which will be applied to harsh radiation environments. In this study, we fabricated the SiC radiation detectors and measured the current-voltage characteristics of SiC detectors irradiated by Co-60 gamma-ray source. The I-V curves showed a decrease of the leakage current with increasing dose rate of Co-60 gamma-ray in the 0-100V bias voltage range.

키워드

SiC radiation detectorradiation damagecurrent-voltage characteristicsgamma-ray irradiationsemiconductor detector
제목
Study of the current-voltage characteristics of a SiC radiation detector irradiated by Co-60 gamma-rays
저자
Kang, Sang MookHa, Jang Ho Park, Se HwanKim, HSChun, Sung DaeKim, Yong Kyun
DOI
10.1016/j.nima.2007.04.025
발행일
2007-08
유형
Article; Proceedings Paper
저널명
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
579
1
페이지
145 ~ 147