In ultrahigh field magnetic resonance imaging (MRI) using a static magnetic field over 7 T, the inhomogeneity in the time-varying magnetic field ( B1), which is caused by its short wavelength, is mitigated by using a multichannel radio frequency (RF) coil as an antenna. However, the conventional multichannel RF coil has a fixed configuration; consequently, its transmit efficiency (TE) changes according to the head shape of the patient. In this study, a geometrically adjustable eight-channel RF transmitter head coil for 7-T MRI was proposed. All resonators and supporting structures (called bridges) in the coil have interlocking teeth at their sides, allowing each channel to adjust its geometric location according to the head shape of the patient. Scenarios for conventional and proposed coils in three different configurations were simulated and compared in terms of the transmitted component of the B1 field ( B+ 1 field) and specific absorption rate (SAR) maps. The simulation results indicate that, for a given configuration and head shape, the TE of the proposed coil was higher than that of the conventional coil, while the SAR efficiency remained unchanged. Furthermore, the S-parameter matrix of the proposed coil and the measured B1 field produced by the two channels facing each other show that the proposed coil resonates at 298 MHz with acceptable mutual coupling and effectively improves the B1 field intensity. Moreover, the obtained signal-to-noise ratio (SNR) maps and magnetic resonance (MR) images prove that the proposed coil is an efficient candidate for an RF transmitter or transceiver head coil used in 7-T MRI.

키워드

Radio frequencyHeadMagnetic resonance imagingShapeMagnetic headsResonatorsShape measurementAdjustable coilmagnetic resonance imaging (MRI)radio frequency (RF) coil7 Ttransmitter head coilSTEPPED IMPEDANCE RESONATORS7 TPHASED-ARRAYDESIGNMRIRECEIVEFIELDS
제목
Adjustable RF Transmitter Head Coil: Improving Transmit Efficiency With SAR Management for 7-T Magnetic Resonance Imaging
저자
Cho, YoungdaeBasir, AbdulYoo, Hyoung suk
DOI
10.1109/TMTT.2021.3057620
발행일
2021-05
유형
Article
저널명
IEEE Transactions on Microwave Theory and Techniques
69
5
페이지
2686 ~ 2696