상세 보기
속도 분포 기반 단층촬영을 위한 최적의 트랜스듀서의 조합
- Kim, Young Hun;
- Park, Kwan Kyu
초록
The type of ultrasound transducer used influences the quality of a reconstructed ultrasound image. This study analyzed the effect of transducer type on ultrasound computed tomography (UCT) image quality. The UCT was modeled in an ultrasound simulator by using a 5 cm anatomy model and a ring-shape 5 MHz 128 transducer array, which considered attenuation, refraction, and reflection. Speed-of-sound images were reconstructed by the Radon transform as the UCT image modality. Acoustic impedance images were also reconstructed by the delay-and-sum (DAS) method, which considered the speed of sound information. To determine the optimal combination of transducers in observation, point-source, flat, and focused transducers were tested in combination as trasmitters and receivers; UCT images were constructed from each combination. The combination of point-source/flat transducer as transmitting and receiving devices presented the best reconstructed image quality. In UCT implementation, the combination of a flat transducer for transmitting and a point transducer for receiving permitted acceptable image quality.
키워드
- 제목
- 속도 분포 기반 단층촬영을 위한 최적의 트랜스듀서의 조합
- 제목 (타언어)
- Transducer Combination for High-Quality Ultrasound Tomography Based on Speed of Sound Imaging
- 저자
- Kim, Young Hun; Park, Kwan Kyu
- 발행일
- 2016-02
- 유형
- Article
- 저널명
- 비파괴검사학회지
- 권
- 36
- 호
- 1
- 페이지
- 27 ~ 34