속도 분포 기반 단층촬영을 위한 최적의 트랜스듀서의 조합

Transducer Combination for High-Quality Ultrasound Tomography Based on Speed of Sound Imaging

초록

The type of ultrasound transducer used influences the quality of a reconstructed ultrasound image. This study analyzed the effect of transducer type on ultrasound computed tomography (UCT) image quality. The UCT was modeled in an ultrasound simulator by using a 5 cm anatomy model and a ring-shape 5 MHz 128 transducer array, which considered attenuation, refraction, and reflection. Speed-of-sound images were reconstructed by the Radon transform as the UCT image modality. Acoustic impedance images were also reconstructed by the delay-and-sum (DAS) method, which considered the speed of sound information. To determine the optimal combination of transducers in observation, point-source, flat, and focused transducers were tested in combination as trasmitters and receivers; UCT images were constructed from each combination. The combination of point-source/flat transducer as transmitting and receiving devices presented the best reconstructed image quality. In UCT implementation, the combination of a flat transducer for transmitting and a point transducer for receiving permitted acceptable image quality.

키워드

UltrasoundComputed Tomography (CT)Radon TransformDelay and SumTransducerTime of FlightSimulationAcoustics초음파단층촬영라돈 변환시간지연트랜스듀서전파시간시뮬레이션음향
제목
속도 분포 기반 단층촬영을 위한 최적의 트랜스듀서의 조합
제목 (타언어)
Transducer Combination for High-Quality Ultrasound Tomography Based on Speed of Sound Imaging
저자
Kim, Young HunPark, Kwan Kyu
DOI
10.7779/JKSNT.2016.36.1.27
발행일
2016-02
유형
Article
저널명
비파괴검사학회지
36
1
페이지
27 ~ 34