상세 보기
Comparison of Fresnel lens-based focused vortex transducer and focused transducer for neurostimulation
- Kim, Young Hun;
- Kang, Ki Chang;
- Park, Kwan Kyu;
- Firouzi, Kamyar;
- Khuri-Yakub, Burtus T.
WEB OF SCIENCE
0SCOPUS
0초록
The most common neurostimulation method is the stimulation of the focal point by using a focused ultrasound transducer. In the field of particle manipulation using ultrasound, research using acoustic vortices is ongoing and presents promising characteristics. In neurostimulation using ultrasound, the acoustic radiation force is a leading candidate for a physical mechanism. In this study, the acoustic radiation forces generated by a general-focused transducer and a vortex transducer were compared. For the experiment, a 70 mm diameter transducer with a center frequency of 1 MHz was fabricated using a piezoceramic plate. The ultrasound transducer frame was made of aluminum, and air backing was used to maximize the output pressure. A large-scale Fresnel focused lens and focused vortex lens were fabricated using polydimethylsiloxane (PDMS). A mold of an 8-step Fresnel lens with a diameter of 70 mm was developed using 3-D printing. We compared the pressure field and radiation force according to the two manufactured lenses to determine whether there was any difference in neurostimulation.
키워드
- 제목
- Comparison of Fresnel lens-based focused vortex transducer and focused transducer for neurostimulation
- 저자
- Kim, Young Hun; Kang, Ki Chang; Park, Kwan Kyu; Firouzi, Kamyar; Khuri-Yakub, Burtus T.
- 발행일
- 2021-11
- 유형
- Proceedings Paper
- 저널명
- INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS 2021)
- 권
- 2021
- 호
- September
- 페이지
- 1 ~ 4