개선된 계단 배기 방법을 이용한 혈압 측정

Blood Pressure Measurement using the Modified Step-wise Deflation Method
  • 오홍식
  • 이종실
  • 지영준
  • 김인영

초록

In the automatic non-invasive blood pressure measurement device, the oscillometric method is widely used. In the oscillometric method, the step-wise deflation has the advantage of the robustness for the motion artifacts than the linear deflation method. But it has the disadvantage of its longer measurement time because we need to detect two or more pulses in a certain cuff pressure step. In this study, we suggest the modified step-wise deflation method to overcome this limitation while maintaining the general concept of step-wise deflation. Using one valid pulse in each step and the deflating valve control during the diastolic period, the measurement time could be reduced. In order to verify the accuracy of the proposed algorithm, we compared the blood pressure values from the suggested method and the blood pressure values from the conventional auscultation method. The mean and standard deviation were -0.50±5.3mmHg and 2.08±4.75mmHg, for systolic and diastolic blood pressure respectively. The measurement time can be reduced up to the half of conventional step-wise deflation method.

키워드

peak detectionoscillometricnon-invasive blood pressurestep deflation
제목
개선된 계단 배기 방법을 이용한 혈압 측정
제목 (타언어)
Blood Pressure Measurement using the Modified Step-wise Deflation Method
저자
오홍식이종실지영준김인영
발행일
2010-10
저널명
의공학회지
31
5
페이지
351 ~ 358