New calculation method for 3D dose distribution in tetrahedral-mesh phantoms in Geant4

  • Han, Min Cheol
  • Ku, Youngmo
  • Lee, Hyun Su
  • Yeom, Yeon Soo
  • Han, Haegin
  • ... Kim, Chan Hyeong
Citations

WEB OF SCIENCE

3
Citations

SCOPUS

4

초록

The tetrahedral-mesh (TM) geometry, which is a very promising geometry for computational human phantoms, has a limitation in 3D dose distribution calculation for medical applications. Even though Geant4 provides the read-out geometry for calculating 3D dose distribution in the TM geometry, this method significantly slows down the computation speed. In the present study, we developed a new method, called Moving Voxel-based Dose-Distribution Calculator (MVDDC), to rapidly calculate a 3D dose distribution in a TM geometry. To evaluate the performance of the MVDDC method, a simple TM cubic phantom and a human phantom were implemented in Geant4. Subsequently, the phantoms were irradiated with proton spot beams under various conditions, and the obtained results were compared with those of the read-out geometry method. The results show that there is no significant difference between the dose distributions calculated using the new method and the read-out geometry method. With respect to the computational performance, the speeds of simulations using the MVDDC were approximately 1.4-2.7 times faster than those of the simulations using the read-out geometry method.

키워드

Tetrahedral-meshComputational phantom3D dose distributionMonte CarloGeant4COMPUTATIONAL PHANTOMSPATIENTMODELS
제목
New calculation method for 3D dose distribution in tetrahedral-mesh phantoms in Geant4
저자
Han, Min CheolKu, YoungmoLee, Hyun SuYeom, Yeon SooHan, HaeginKim, Chan Hyeong
DOI
10.1016/j.ejmp.2019.09.239
발행일
2019-10
유형
Article
저널명
Physica Medica
66
페이지
97 ~ 103