Development of a sampling method for carbonyl compounds released due to the use of electronic cigarettes and quantitation of their conversion from liquid to aerosol

Citations

WEB OF SCIENCE

37
Citations

SCOPUS

42

초록

In this study, an experimental method for the collection and analysis of carbonyl compounds (CCs) released due to the use of electronic cigarettes (e-cigarettes or ECs) was developed and validated through a series of laboratory experiments. As part of this work, the conversion of CCs from a refill solution (e-solution) to aerosol also was investigated based on mass change tracking (MCT) approach. Aerosol samples generated from an e-cigarette were collected manually using 2,4-dinitrophenylhydrazine (DNPH) cartridges at a constant sampling (puffing) velocity of 1 L min(-1) with the following puff conditions: puff duration (2 s), interpuff interval (10 s), and puff number (5,10, and 15 times). The MCT approach allowed us to improve the sampling of CCs through critical evaluation of the puff conditions in relation to the consumed quantities of refill solution. The emission concentrations of CCs remained constant when e-cigarettes were sampled at or above 10 puff. Upon aerosolization, the concentrations of formaldehyde and acetaldehyde increased 6.23- and 58.4-fold, respectively, relative to their concentrations in e-solution. Furthermore, a number of CCs were found to be present in the aerosol samples which were not detected in the initial e-solution (e.g., acetone, butyraldehyde, and o-tolualdehyde).

키워드

Electronic cigarettesStandard operating procedureAerosolDNPH cartridgeFormaldehydeREPLACEMENT LIQUIDSMASS-SPECTROMETRYCHROMATOGRAPHYCARTRIDGESNICOTINEVAPORS
제목
Development of a sampling method for carbonyl compounds released due to the use of electronic cigarettes and quantitation of their conversion from liquid to aerosol
저자
Jo, Sang-HeeKim, Ki-Hyun
DOI
10.1016/j.chroma.2015.12.061
발행일
2016-01
유형
Article
저널명
Journal of Chromatography A
1429
페이지
369 ~ 373