상세 보기
Development and performance evaluation of large-area hybrid gamma imager (LAHGI)
- Lee, Hyun Su;
- Kim, Jae Hyeon;
- Lee, Junyoung;
- Kim, Chan Hyeong
WEB OF SCIENCE
5SCOPUS
8초록
We report the development of a gamma-ray imaging device, named Large-Area Hybrid Gamma Imager (LAHGI), featuring high imaging sensitivity and good imaging resolution over a broad energy range. A hybrid collimation method, which combines mechanical and electronic collimation, is employed for a stable imaging performance based on large-area scintillation detectors for high imaging sensitivity. The system comprises two monolithic position-sensitive NaI(Tl) scintillation detectors with a crystal area of 27 × 27 cm2 and a tungsten coded aperture mask with a modified uniformly redundant array (MURA) pattern. The performance of the system was evaluated under several source conditions. The system showed good imaging resolution (i.e., 6.0–8.9° FWHM) for the entire energy range of 59.5–1330 keV considered in the present study. It also showed very high imaging sensitivity, successfully imaging a 253 μCi 137Cs source located 15 m away in 1 min; this performance is notable considering that the dose rate at the front surface of the system, due to the existence of the 137Cs source, was only 0.003 μSv/h, which corresponds to ∼3% of the background level.
키워드
- 제목
- Development and performance evaluation of large-area hybrid gamma imager (LAHGI)
- 저자
- Lee, Hyun Su; Kim, Jae Hyeon; Lee, Junyoung; Kim, Chan Hyeong
- 발행일
- 2021-02
- 유형
- Article in Press
- 권
- 53
- 호
- 8
- 페이지
- 2640 ~ 2645