Prompt gamma measurements for locating the dose falloff region in the proton therapy

Citations

WEB OF SCIENCE

344
Citations

SCOPUS

426

초록

The location of the distal falloff in the proton therapy is an important but often uncertain parameter as different tissue elements are traversed by the beam. A multilayered collimator system has been constructed as a practical means to locate the dose ends by measuring prompt gammas. The collimator is designed to moderate and capture fast neutrons and to prevent unwanted gammas from reaching the scintillation detector. The system has been studied using Monte Carlo technique and has been tested in the beam energy range of 100-200 MeV. Measurements clearly indicated correlations between the gamma distributions and the distal falloff regions.

키워드

CANCERPETRADIATIONBEAMS
제목
Prompt gamma measurements for locating the dose falloff region in the proton therapy
저자
Min, Chul-HeeKim, Chan HyeongYoun, Min-YoungKim, Jong-Won
DOI
10.1063/1.2378561
발행일
2006-10
유형
Article
저널명
Applied Physics Letters
89
18
페이지
183517(1) ~ 183517(3)