Radiation Hardness Test of a Silicon Detector under Radiation Dose Rate of Nuclear Power Plant for In-Containment Coolant Leakage Detection System

Citations

WEB OF SCIENCE

2

초록

An influence evaluation by background radiation on a silicon detector, which will be used to detect a coolant leakage, and installed in a containment building of a nuclear power plant, was performed and the result was discussed. The detector that consists of a silicon sensor and preamplifier mounted in a shielding structure which composed of a 5 cm lead cylinder will be installed in an annulus zone that is influenced by background radiation (neutron and gamma ray) from an operation of a nuclear reactor. Absorbed dose rates on a silicon sensor and preamplifier were calculated as 2.15 mGy/hr and 1.05 mGy/hr, respectively, by Monte Carlo N-Particle (MCNP) simulation. Data of background radiation had referred to a Final Safety Analysis Report (FSAR) of a nuclear power plant in the Republic of Korea. A silicon sensor and preamplifier were irradiated by a Co-60 gamma radiation source equipped in a facility of Korea Atomic Energy Research Institute Advanced Radiation Technology Institute (KAERI ARTI) of the Republic of Korea. A Po-210 alpha source was used as a check source to evaluate a state of a function of the detector during gamma irradiation. Absorbed dose rates were about 22.92 mGy/hr and 6.6 mGy/hr on silicon sensor and preamplifier, respectively. Before and during gamma irradiation, a count rates from the check source wasn't changed (from 18.4 cps to 18.4 +/- 0.2 cps after irradiation), and any degradations of function also weren't observed. Even more harsh condition than calculated dose rates referred by the condition of background radiation of in-containment, the silicon detector maintained the ability of function of charged particles detection. Based on the result, it has been demonstrated that a silicon detector is a suitable detector for detecting charged particles from a leaked coolant even during interfered by the background radiation of a primary system of a nuclear power plant.

키워드

Silicon detectorIrradiation testCoolant leakage detection system
제목
Radiation Hardness Test of a Silicon Detector under Radiation Dose Rate of Nuclear Power Plant for In-Containment Coolant Leakage Detection System
저자
Kim, JongheonPak, KihongPark, JunesicKim, Yong Kyun
DOI
10.1051/epjconf/202125305004
발행일
2021-06
유형
Proceedings Paper
저널명
ANIMMA 2021 - ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS
253
페이지
1 ~ 4

파일 다운로드