상세 보기
초록
Purpose: To compare central corneal thickness (CCT) as measured using noncontact specular microscopy (NCSM), dual rotating Scheimpflug camera (Galilei®), and ultrasound pachymetry (USP). Methods: The measurements of CCT using NCSM, dual rotating Scheimpflug camera and USP in 70 eyes of 70 healthy subjects were compared. Results: The average measurements of CCT using NCSM, dual rotating Scheimpflug camera, and USP were 567.70 ± 31.21 μm, 557.84 ± 26.29 μm, and 553.31 ± 29.69 μm, respectively. The CCT measurement using NCSM was statistically significantly thicker than when measured using USP (p < 0.05). There was no significant difference between the NCSM and dual rotating Scheimpflug camera (p = 0.138). Additionally, there was no significant difference between the dual rotating Scheimpflug camera and USP (p = 0.656). A significant linear correlation was observed among the NCSM, dual rotating Scheimpflug camera, and USP (r > 0.900, p < 0.001). Conclusions: The results of the 3 methods were significantly correlated but the measurement using NCSM was significantly thicker than when using USP. CCT measurements of healthy eyes using dual rotating Scheimpflug camera were more correlated with USP than NCSM. The CCT measurements using dual rotating Scheimpflug camera is a better alternative for USP than NCSM.
키워드
- 제목
- 비접촉경면현미경, 이중샤임플러그 전안부사진기, 초음파각막두께측정계를 이용해 측정한 중심각막두께의 비교
- 제목 (타언어)
- Central Corneal Thickness Measured by Noncontact Specular Microscopy, Dual Rotating Scheimpflug Camera and Ultrasound Pachymetry
- 저자
- 이민지; 신용운; 임한웅; 강민호; 조희윤; 성민철
- 발행일
- 2015-10
- 저널명
- 대한안과학회지
- 권
- 56
- 호
- 10
- 페이지
- 1520 ~ 1526