상세 보기
Design of a Loop Resonator with a Split-Ring-Resonator (SRR) for a Human-body Coil in 3 T MRI Systems
- Son, Hyeok Woo;
- Cho, Young Ki;
- Kim, Byung Mun;
- Back, Hyun Man;
- Yoo, Hyoung suk
WEB OF SCIENCE
2SCOPUS
3초록
A new radio-frequency (RF) resonator for Nuclear Magnetic Resonance (NMR) imaging at clinical magnetic resonance imaging (MRI) systems is proposed in this paper. An approach based on the effects of the properties of metamaterials in split-ring resonators (SRRs) is used to design a new loop resonator with a SRR for NMR imaging. This loop resonator with a SRR is designed for NMR imaging at 3 T MRI systems. The 3D electromagnetic simulation was used to optimize the design of the proposed RF resonator and analyze it’s performance at 3 T MRI systems. The proposed RF resonator provides strong penetrating magnetic fields at the center of the human phantom model, approximately 10%, as compared to the traditional loop-type RF resonator used for NMR imaging at clinical MRI systems. We also designed an 8-channel body coil for human-body NMR imaging by using the proposed loop resonator with a SRR. This body coil also produces more homogeneous and highly penetrating magnetic fields into the human phantom model.
키워드
- 제목
- Design of a Loop Resonator with a Split-Ring-Resonator (SRR) for a Human-body Coil in 3 T MRI Systems
- 저자
- Son, Hyeok Woo; Cho, Young Ki; Kim, Byung Mun; Back, Hyun Man; Yoo, Hyoung suk
- 발행일
- 2016-04
- 유형
- 정기학술지(Article(Perspective Article포함))
- 권
- 68
- 호
- 7
- 페이지
- 908 ~ 913
- 언어
- ENG
- 출판사
- 한국물리학회
- 발행국가
- 대한민국
- 분량
- 6 페이지
- ISSN
- E 1976-8524
P 0374-4884