Mass Background Spectra of ICP-MS with Various Acids

Citations

WEB OF SCIENCE

5
Citations

SCOPUS

5

초록

Background spectra of ICP-MS performed with acid can interfere with the spectra of sample analytes. Thus, we have investigated the background spectra of various acids for use in elemental analysis by argon inductively coupled plasma mass spectrometry. The background spectra for 2% HNO3, 5% HNO3, 5% H2SO4, and 4% CH3COOH were obtained with ICP-MS. This study also shows the possibility of the interference of Na, which could not be from the matrix, but from the ICP-MS sampler. The source of interference of boron was the vibrating glass surface of ultrasonic nebulizer. After a mug cup was kept in 4% CH3COOH for 24 hours, trace Pb was determined with ICP-MS.

키워드

ICP-MSAcidsBackgroundSpectraINDUCTIVELY-COUPLED PLASMASSPECTROMETRY
제목
Mass Background Spectra of ICP-MS with Various Acids
저자
Nam, Sang-HoChung, HoeilKim, Jae-JinLee, Yong-Ill
DOI
10.5012/bkcs.2008.29.11.2237
발행일
2008-11
유형
Article
저널명
Bulletin of the Korean Chemical Society
29
11
페이지
2237 ~ 2240

파일 다운로드