Study of n/gamma discrimination using He-3 proportional chamber in high gamma-ray fields

Citations

WEB OF SCIENCE

5
Citations

SCOPUS

6

초록

The He-3 proportional chamber is widely used for neutron measurement owing to its high neutron detection efficiency and simplicity for gamma-ray rejection. In general, the neutron and gamma-ray signals obtained from the He-3 proportional chamber can be easily separated by the difference in the pulse heights. However, for a high gamma-ray field, the gamma-ray signal cannot be precisely eliminated by the pulse height due to gamma-ray pulse pileup which causes the pulse height of gamma-ray pulse to increase and making the pulses due to neutrons and gamma rays indistinguishable. In this study, an improved algorithm for n/gamma discrimination using a parameter, which is the ratio of the rise time to the pulse height, is proposed. The n/gamma discrimination performance of the algorithm is evaluated by applying it to Cf-252 neutron signal separation from various gamma-ray exposure rate levels ranging 0.1-5 R/h. The performance is compared to that of the conventional pulse-height analysis method in terms of the gamma elimination ratio. The suggested algorithm shows better performance than the conventional one by 1.7% (at 0.1 R/h) to 70% (at 5 R/h) for gamma elimination.

키워드

He-3 proportional chamberNeutron detectorn/gamma discriminationAlgorithmHigh gamma-ray fieldCOUNTERS
제목
Study of n/gamma discrimination using He-3 proportional chamber in high gamma-ray fields
저자
Choi, JoonbumPark, JunesicSon, JaebumKim, Yong Kyun
DOI
10.1016/j.net.2018.08.013
발행일
2019-02
유형
Article
저널명
Nuclear Engineering and Technology
51
1
페이지
263 ~ 268

파일 다운로드