Low-Dose Computed Tomography of the Chest Using Iterative Reconstruction Versus Filtered Back Projection: Comparison of Image Quality

Citations

WEB OF SCIENCE

27
Citations

SCOPUS

25

초록

Objective: The purpose of this study was to compare image quality of iterative reconstruction (IR) to filtered back projection (FBP) in low-dose computed tomography of the chest. Methods: Forty-three consecutive patients were retrospectively enrolled. Eight series of images were reconstructed using FBP and 7 levels of IR in each subject. Image noise, signal-to-noise ratio (SNR), and SNR improvement were measured. Two radiologists evaluated subjective artifact, image artificiality, and subjective overall image quality with 4- or 5-point scales. Results: Iterative reconstruction showed significantly lower image noise (135.5 +/- 36.6 vs 219.9 +/- 40.9) and higher SNR (0.36 +/- 0.12 vs 0.21 +/- 0.05) than FBP (P < 0.001). Signal-to-noise ratio improvement was 72.4% +/- 44.9%. Subjective artifact of FBP was significantly higher than IR images (P < 0.001). Image artificiality of IR was significantly higher than that of FBP (P < 0.001). Overall, subjective image quality was poor in FBP and acceptable or good in IR. Conclusions: With the use of IR, low-dose computed tomography of the chest would achieve less image noise and better image quality compared to the FBP.

키워드

lung; computed tomography; iterative reconstruction; radiation dose; image quality; CT; REDUCTION
제목
Low-Dose Computed Tomography of the Chest Using Iterative Reconstruction Versus Filtered Back Projection: Comparison of Image Quality
저자
Lee, Youkyung; Jin, Kwang Nam; Lee, Nyoung Keun
DOI
10.1097/RCT.0b013e318260ce39
발행일
2012-09
유형
Article
저널명
Journal of Computer Assisted Tomography
권
36
호
5
페이지
512 ~ 517